User Manual
™
Perfecto Series
2
Oxygen Concentrators without SensO ®,
2
Oxygen Concentrators with SensO
2
®
HomeFill Compatible
Perfecto with SensO Perfecto
2 2 2
Model IRC5PO2 Model IRC5P
Model IRC5PO2W Model IRC5PW
Model IRC5PO2V
Dealer: This manual MUST be given to the end
user of this product.
User: BEFORE using this product, read this
manual and save for future reference.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
Résumé du contenu de la page N° 2
WARNING WARNING DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UNDERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS OR INSTRUCTIONS, CONTACT A HEALTHCARE PROFESSIONAL, DEALER OR TECHNICAL PERSONNEL BEFORE ATTEMPTING TO USE THIS EQUIPMENT - OTHERW
Résumé du contenu de la page N° 3
Table of Contents TABLE OF CONTENTS ACCESSORIES ..................................................2 SPECIAL NOTES ...............................................4 LABEL LOCATION ............................................6 SECTION 1—GENERAL GUIDELINES ..................7 To Reduce the Risk of Burns, Electrocution, Fire or Injury to Persons ......................................................................................................... 7 Operating Information ..........................
Résumé du contenu de la page N° 4
SPECIAL NOTES SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words. SIGNAL WORD MEANING Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious DANGER injury. Warning indicates a potentially hazardous situation which, if not avoided, could result in death or WARNING serious injury
Résumé du contenu de la page N° 5
SPECIAL NOTES CAUTION “Caution: Federal law restricts this device to sale or rental by or on order of a physician, or any other practitioner licensed by the law of the State in which he/she practices to use or order the use of this device.” Invacare recommends an alternate source of supplemental oxygen in the event of a power outage, alarm condition or mechanical failure. Consult your physician or equipment provider for the type of reserve system required. This equipment is to be used as a
Résumé du contenu de la page N° 6
P / N 1 1 4 7 9 2 0 R e v . B LABEL LOCATION LABEL LOCATION Perfecto ™ 2 FOR OPERATING INSTRUCTIONS or ALARMS (See Operator’s Manual) DANGER - RISK OF FIRE NO SMOKING - Keep ALL sources of ignition out of the room in which this product is located and away from areas where oxygen is being delivered. Textiles, oil and other combustibles are easily ignited and burn with great intensity in oxygen enriched air. DANGER - ELECTRIC SHOCK DO NOT remove cover. Refer servicing to qualif
Résumé du contenu de la page N° 7
SECTION 1—GENERAL GUIDELINES SECTION 1—GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the Perfecto concentrator these instructions 2 MUST be followed. WARNING This section contains important information for the safe operation and use of this product. To Reduce the Risk of Burns, Electrocution, Fire or Injury to Persons DANGER Risk of electric shock. DO NOT disassemble. Refer servicing to qualified service personnel. No user serviceable parts. Avo
Résumé du contenu de la page N° 8
SECTION 1—GENERAL GUIDELINES Operating Information For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter periods of operation may reduce maximum product life. Keep the oxygen tubing, cord, and unit out from under such items as blankets, bed coverings, chair cushions, clothing and from heated or hot surfaces, including space heaters, away stoves and similar electrical appliances. DO NOT move or relocate concen
Résumé du contenu de la page N° 9
SECTION 1—GENERAL GUIDELINES Other devices may experience interference from even the low levels of electromagnetic emissions permitted by the above standards. To determine if the emissions from the Perfecto is 2 causing the interference, turn the Perfecto Off. If the 2 interference with the other device(s) stops, then the Perfecto 2 is causing the interference. In such rare cases, interference may be reduced or corrected by one of the following measures: •Reposition, relocate, or incre
Résumé du contenu de la page N° 10
SECTION 2—FEATURES SECTION 2—FEATURES Oxygen Outlet Oxygen Purity Indicator Lights / Fault and Power Indicator Lights Flowmeter Circuit Breaker Elapsed Time Power Switch Meter ACCESSORIES (NOT SHOWN): PreciseRX™ Pediatric Humidifier/Flowmeter Accessory - IRCPF16 (Perfecto 5 Only) 2 HomeFill home oxygen compressor - IOH200 REAR VIEW *NOTE: This outlet fitting is to be Cabinet Filter used only for filling oxygen cylinders with the HomeFill home oxygen compressor. The outlet fitting does not
Résumé du contenu de la page N° 11
SECTION 3—HANDLING SECTION 3—HANDLING The concentrator should ALWAYS be kept in the upright position to prevent cabinet damage while being transported. If the concentrator is to be reshipped by common carrier, additional cartons are available from Invacare. Unpacking NOTE: For this procedure, refer to FIGURE 3.1. 1. Check for any obvious damage to the carton or its contents. If damage is evident, notify the carrier, or your local dealer. 2. Remove all loose packing from the carton. 3. C
Résumé du contenu de la page N° 12
SECTION 4—TYPICAL PRODUCT PARAMETERS SECTION 4—TYPICAL PRODUCT PARAMETERS Electrical Requirements: 120 VAC ± 10% (132 VAC/108 VAC), 60 Hz Rated Current Input: 3 A Sound Level: 43 ± 2 dBA Average for Perfecto Models and 2 Perfecto V Model 2 40 ± 2 dBA Average for Perfecto W Models 2 Altitude: All Perfecto Models: Up to 8,000 ft (2438 2 meters) above sea level without degredation of concentration levels. From 8,000 ft to 13,129 (4000 meters) below 90% efficiency. All Perfecto Models while fi
Résumé du contenu de la page N° 13
SECTION 4—TYPICAL PRODUCT PARAMETERS Change in maximum 0.7 L/min recommended flow when back pressure of 7kPa is applied: Filters: Cabinet, Outlet HEPA and Compressor Inlet Safety System: Current overload or line surge shutdown. High temperature compressor shutdown. High Pressure Alarm w/compressor shutdown. Low Pressure Alarm w/compressor shutdown. Battery Free Power Loss Alarm. SensO Oxygen 2 System (SensO Model) Low Flow Alarm. 2 Width: 15 in ± 3/8 in (38.1 cm ± 1 cm) Height: 23 in ± 3/
Résumé du contenu de la page N° 14
SECTION 4—TYPICAL PRODUCT PARAMETERS Placement: No closer than three inches from any wall, furniture, draperies, or similar surfaces. Tubing: 7 ft cannula with a maximum 50 ft of Crush-Proof Tubing (DO NOT pinch). Relative Humidity: 20 to 60% Time of Operation: Up to 24 hours per day. Recommended Storage -20°F to 150°F (-29°C to 65°C), at 15-95% and Shipping Tempera- relative humidity. ture: *NOTE: Stated concentration levels achieved after initial warm‐up period (approximately 30 minutes).
Résumé du contenu de la page N° 15
SECTION 5—OPERATING INSTRUCTIONS SECTION 5—OPERATING INSTRUCTIONS Introduction Your oxygen concentrator is intended for individual use indoors. It is an electronically operated device that separates oxygen from room air. It provides high concentration of oxygen directly to you through a nasal cannula. Clinical studies have documented that oxygen concentrators are therapeutically equivalent to other types of oxygen delivery systems. Your provider will show you how to use your oxygen concen
Résumé du contenu de la page N° 16
SECTION 5—OPERATING INSTRUCTIONS Set Up Plug in Power Cord Plug in power cord to an electrical outlet. Connect Humidifier (If So Prescribed) WARNING DO NOT overfill humidifier. DO NOT reverse the oxygen input and output connections. Water from the humidifier bottle will travel through the cannula back to the patient. NOTE: For this procedure, refer to FIGURE 5.1, FIGURE 5.2 on page17, FIGURE 5.3 on page18. 1. Remove cap from bottle. 2. Fill humidifier with distilled water to the level indic
Résumé du contenu de la page N° 17
SECTION 5—OPERATING INSTRUCTIONS Humidifier Bottle Adapter Insert Flat Head Screwdriver Here Cabinet Filter Filter Access Door FIGURE 5.2 Humidifer Bottle Adapter 5. Attach the humidifier bottle adapter to the humidifier bottle by turning the wing nut on the humidifier bottle counterclockwise until it is securely fastened. See Detail “A” in FIGURE 5.3. 6. Place the humidifier bottle/adapter assembly in the humidifier compartment on the concentrator. See Detail “B” in FIGURE 5.3. 7. At
Résumé du contenu de la page N° 18
SECTION 5—OPERATING INSTRUCTIONS Oxygen Outlet DETAIL “A” DETAIL “B” Connector Oxygen Humidifier Tubing Bottle Humidifier Adapter Bottle Outlet Humidifier Bottle Humidifier Bottle Humidifier Compartment FIGURE 5.3 Humidifier Compartment Power Switch NOTE: For this procedure, refer to FIGURE 5.4. On/Off (I/O) 1. Press power switch to on Power position. All the panel Switch lights and the audible alarm will come on for one second, indicating that the unit is functioning prope
Résumé du contenu de la page N° 19
SECTION 5—OPERATING INSTRUCTIONS Flowrate NOTE: For this procedure, refer to FIGURE 5.5 on page20. 1. Turn the flowrate knob to the setting prescribed by your physician or therapist. CAUTION Never set the flow greater than 5 L/min. WARNING DO NOT change the L/min setting on the flowmeter unless a change has been prescribed by your physician or therapist. NOTE: To properly read the flowmeter, locate the prescribed flowrate line on the flowmeter. Next, turn the flow knob until the ball rises
Résumé du contenu de la page N° 20
SECTION 5—OPERATING INSTRUCTIONS Flow Knob Flowmeter 0.5 L/min Ball Front Panel FIGURE 5.5 Flowrate Front Panel Indicators - Units with SensO 2 The SensO feature monitors the purity of oxygen generated 2 by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate. OXYGEN PURITY Auto Shut Down RED Use Backup Call Supplier Below Normal YELLOW Normal GREEN O 2 FIGURE 5.6 Front Panel Indicators - Units with SensO2 ™ P