Owner’s Operator and Maintenance Manual
MA80 Series
MA80 True Low Air Loss
System
MA85 Five-Zone Alternating
Pressure and True Low Air
Loss System
DEALER: This manual MUST be given to
the user of the product.
USER: BEFORE using this product, read this
manual and save for future reference.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
Résumé du contenu de la page N° 2
SPECIAL NOTES WARNING DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied with this product or optional equipment. If you are unable to understand the warnings, cautions or instructions, contact a healthcare professional, dealer or technical personnel before attempting to use this equipment - otherw
Résumé du contenu de la page N° 3
TABLE OF CONTENTS TABLE OF CONTENTS SPECIAL NOTES ................................................................................2 REGISTER YOUR PRODUCT ............................................................... 4 TYPICAL PRODUCT PARAMETERS .................................................... 5 Electrical Parameters.............................................................................................................................5 Performance Parameters............................
Résumé du contenu de la page N° 4
TABLE OF CONTENTS TABLE OF CONTENTS Transferring Patient From/To a Gurney .............................................................................................19 Transferring Patient From/To a Wheelchair......................................................................................19 Preparing for CPR Procedure................................................................................................................19 SECTION 3—MAINTENANCE AND TROUBLESHOOTING ................
Résumé du contenu de la page N° 5
TYPICAL PRODUCT PARAMETERS TYPICAL PRODUCT PARAMETERS Electrical Parameters MA80/MA85 INPUT VOLTAGE AC: 90 V INPUT FREQUENCY: 60 Hz CURRENT: 2 A MAXIMUM POWER CONSUMPTION: 150 ± 10 W CIRCUIT PROTECTION: Dual fused, 250 V, 5 A fast blow fuses MODE OF OPERATION: Continuous Performance Parameters MA80 MA85 WEIGHT CAPACITY STANDARD MATTRESS: 350 lbs BARIATRIC MATTRESS: 1000 lbs PRESSURE ZONE: 5 MAXIMUM FLOW: 1275 ± 100 LPM MAXIMUM PRESSURE: 35 ± 5 mmHg MAXIMUM FLOW TIMER: 15 minutes SUPPORT SURFAC
Résumé du contenu de la page N° 6
TYPICAL PRODUCT PARAMETERS Environmental Parameters MA80/MA85 OPERATING CONDITIONS AMBIENT TEMPERATURE: 50° - 104° F RELATIVE HUMIDITY: 30% - 75% Non-Condensing ATMOSPHERIC PRESSURE: 70 - 106 kPa STORAGE AND SHIPPING CONDITIONS AMBIENT TEMPERATURE: -40° - 158° F RELATIVE HUMIDITY: 10% - 100% ATMOSPHERIC PRESSURE: 50 - 106 kPa MA80 Series 6 Part No 1148138
Résumé du contenu de la page N° 7
LABEL LOCATION LABEL LOCATION DANGER-EXPLOSION HAZARD: DO NOT use in the presence of flammable anesthetics. CAUTION: Equipment should be connected to a properly grounded receptacle (3-prong). Risk of Electrical shock. DO NOT remove back. Disconnect air hose before administering CPR. NOTE: This label is on the back of the unit and also contains the ATTENTION serial number. Clean Filter Every 5 months and Whenever Dirty. Remove Filter, Wash, Dry & Replace. WARNING Patient entrapment wit
Résumé du contenu de la page N° 8
SECTION 1—GENERAL GUIDELINES SECTION 1—GENERAL GUIDELINES WARNING SECTION 1 - GENERAL GUIDELINES contains important information for the safe operation and use of this product. Contraindications ALWAYS consult the patient’s physician before using the MA80 and MA85 systems. Installation The MA80 and MA85 systems MUST be installed on medical bed frames with side rails. The side rails MUST be in the raised position whenever a patient is on the bed. Controls on the footboard may be obstructed by
Résumé du contenu de la page N° 9
SECTION 1—GENERAL GUIDELINES Fire Hazard DANGER Smoking DO NOT SMOKE while using this device. This system uses room air for circulation through the mattress. A cigarette can burn a hole in the bed surface and cause damage to the mattress. Also, patient clothing, bed sheets, etc. may be combustible and cause a fire. Failure to observe this warning can result in severe fire, property damage and cause physical injury or death. Smoking by visitors in the room will contaminate the system. There
Résumé du contenu de la page N° 10
SECTION 2—OPERATION SECTION 2—OPERATION CAUTION The control unit and mattress on the MA80 series are designed to be used as a system. DO NOT replace mattresses or control units with other models or other brands. Otherwise, damage to the system may occur. Contact your supplier to get the correct replacement if needed. Installing the Mattress Replacement System CAUTION DO NOT strap the mattress to the bed frame at the head and foot ends. Secure ALL mattress straps. Secure the straps to the bed
Résumé du contenu de la page N° 11
SECTION 2—OPERATION Head End Powered Mattress Buckle Straps Hose Foot End FIGURE 2.1 Installing the Mattress Replacement System Installing the Side Rails WARNING Patient entrapment with bed side rails may cause injury or death. Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment. Follow the manufacturer’s instructions. Monitor patient frequently. Read and understand the Owner’s/Operator’s Manual prior to using this equipment. Invacare product manuals are avai
Résumé du contenu de la page N° 12
SECTION 2—OPERATION Connecting the Hose CAUTION Ensure that the hose connecting the control unit to the mattress is routed such that it cannot be stepped on, kinked, squeezed or otherwise damaged. NOTE: For this procedure, refer to FIGURE 2.3. 1. Locate the hose at the foot end of the mattress. 2. Locate the control unit connector on the right side of the control unit. 3. Lift and hold the connector cover up. 4. Push the hose connector (Detail “A”) onto the control connector. 5. Ensure the hose
Résumé du contenu de la page N° 13
SECTION 2—OPERATION Connecting the Power Cord WARNING DO NOT alter plug to fit a non-conforming outlet. Instead, have an electrician install a properly grounded outlet. Failure to use the correct plug and outlet can result in a potential safety hazard. CAUTION Ensure that the power cord of the control unit is not pinched, or has any objects placed on it, and also ensure it is not located where it can be stepped on or tripped over. NOTE: For this procedure, refer to FIGURE 2.4. 1. Examine t
Résumé du contenu de la page N° 14
SECTION 2—OPERATION Using the Front Panel NOTE: For this procedure, refer to FIGURE 2.5. DETAIL “A” - MA80 FRONT PANEL Low Pressure Fowler Power LED Power Fail Button Pulse Button Button LED Lock/Alarm Static Button Max Inflate Soft Button Button LCD Button Firm Button Display DETAIL “B” - MA85 FRONT PANEL Low Pressure Fowler Power Fail Power Alternating LED Button LED Button Pressure Button Lock/Alarm Static Button Max Inflate Button Soft Button LCD Firm Button Button Display FIGU
Résumé du contenu de la page N° 15
SECTION 2—OPERATION Pulse Button (MA80) 1. Press the Pulse button ( ) to Turn on Pulse mode and select one of four Pulse Times - 1, 3, 5 or 10 minutes. NOTE: The pulse times are 1, 3, 5 or 10 minutes. The pulse times indicate the time between each low pulse. For example, if 5 minutes is selected, all the air cushions in the mattress will inflate for 5 minutes (the selected time) to the selected comfort setting and then drop 20% pressure for 20 seconds. Alternating Pressure Button (MA85) 1.
Résumé du contenu de la page N° 16
SECTION 2—OPERATION Static Button 1. Press the Static button ( ) to select the static mode and maintain the mattress at constant pressure. NOTE: The Static LED is illuminated GREEN. Firm/Soft Buttons 1. Select comfort pressure settings by pressing Firm ( ) or Soft ( ) buttons. • Soft Button - Pressing this button reduces the pressure in the mattress. • Firm Button - Pressing this button increases the pressure in the mattress. NOTE: The patient comfort pressure ranges from Soft (level 0 = 6±4 mm
Résumé du contenu de la page N° 17
SECTION 2—OPERATION 5. Perform one of the following: • Cancel - Press the Fowler button to exit Fowler Transmitter Setup without linking to a transmitter. • Save - Wait 60 seconds without pressing a key. Max Inflate Button 1. Press the Max Inflate button ( ) to select the max inflate mode and inflate the mattress rapidly to maximum firmness (pressurized to 35 ± 6 mmHg). A series of beeps sound every three minutes as a reminder that the Max Inflate mode is active. After 30 minutes, the Max I
Résumé du contenu de la page N° 18
SECTION 2—OPERATION Powering Up the System NOTE: For this procedure, refer to FIGURE 2.5 on page 14. 1. Turn on the power to the system by pressing the Power button on the control unit. NOTE: Once the button is released, a GREEN LED illuminates when the unit is on. Placing the Patient on the Mattress 1. Press the Max Inflate button to turn on the control unit to maximum flow. NOTE: In this mode, the Max Inflate LED lights up. 2. After the mattress is fully inflated, place the patient on the bed
Résumé du contenu de la page N° 19
SECTION 2—OPERATION Transferring Patient From/To a Gurney WARNING ALWAYS engage the wheel locks of the bed and the wheel locks of the gurney before transferring the patient between the bed and the gurney. 1. Engage the wheel locks of the bed. Refer to the owner’s manual provided with the bed. 2. Engage the wheel locks of the gurney. Refer to the owner’s manual provided with the gurney. 3. Press the Max Inflate button to achieve maximum mattress pressure. 4. Raise or lower the bed to match th
Résumé du contenu de la page N° 20
SECTION 2—OPERATION DETAIL “A” Connector Cover Control Unit Connector Control Unit DETAIL “B” Hose Connector Holes FIGURE 2.6 Preparing for CPR Procedure MA80 Series 20 Part No 1148138