Résumé du contenu de la page N° 1
Operation Manual
Model 100E
UV Fluorescence SO Analyzer
2
Also supports operation of: when used in conjunction with:
M100EU Analyzer M100EU Addendum, PN05831
M100EH Analyzer M100EH Addendum, PN 04621
M108E Analyzer M108E Addendum, PN 03940
M108EU Analyzer M108E Addendum, PN03949 and
M100EU Addendum, PN 04621
© TELEDYNE ADVANCED POLLUTION INSTRUMENTATION
9480 CARROLL PARK DRIVE
SAN DIEGO, CA 92121-5201
USA
Toll-free Phone: 800-324-5190
Phone: 858-657-9800
Fax:
Résumé du contenu de la page N° 2
Error! Unknown document property name. Error! Unknown document property name.
Résumé du contenu de la page N° 3
ABOUT THIS MANUAL This M100E manual, PN 04515, is comprised of the following documents: Part No. Rev Name/Description 04515 F M100E Operation Manual (main content) 05036 E Appendix A, Menu Trees and related software documentation, Rev G.4. 05360 Q Spare Parts List (in Appendix B of this manual) 04526 S Recommended Spare Parts Stocking Levels (in Appendix B of this manual) 04728 A AKIT, Spares (in Appendix B of this manual) 04357 A AKIT, Expendables (in Appendix B of this manual) 01475
Résumé du contenu de la page N° 4
Table of Contents Teledyne API M100E Analyzer Operation Manual REVISION HISTORY M100E Operation Manual, PN04515 Document PN RevDCN Change Summary 2011 April 06, Rev F, DCN6048 M100E Operation 04515 F 6048 Administrative corrections in Revision History Manual section: typos. 2011 March 11, Rev E, DCN6019 Combine with M100E “Text” (PN05138) for single M100E Operation 04515 E 6019 document/PN; obsolete 05138. Manual Change Disk-on-Module replacement procedure from A-Crosser to ICOP
Résumé du contenu de la page N° 5
Teledyne API M100E Analyzer Operation Manual Table of Contents Document PN RevDCN Change Summary SPL 05360 Q RSSL 04526 S AKIT, Spares 04728 A AKIT, Expendables 04357 A AKIT, Expendables, IZS 01475 A Repair Request (AppxC) 04796 D Interconnect List 0401101 A Interconnect Diagram 04011 A PCA04003PressFlwSnsr 04354 D PCB,Relay 04524 D PCA05702MoBrd 05703 A PCA04180PMT Preamp 04181 H PCA04258Kybrd Int 04259 A PCA05063 UVDetect 05064 C PCA04692 LmpDrvr 046
Résumé du contenu de la page N° 6
Table of Contents Teledyne API M100E Analyzer Operation Manual This page intentionally left blank. iv 04515F DCN6048
Résumé du contenu de la page N° 7
TABLE OF CONTENTS SECTION I GENERAL INFORMATION .................................................................................. 13 1. INTRODUCTION ................................................................................................................ 15 1.1. Safety Messages...........................................................................................................................................15 1.2. M100E Overview...............................................
Résumé du contenu de la page N° 8
Table of Contents Teledyne API M100E Analyzer Operation Manual 5.12. Special Features .........................................................................................................................................69 SECTION II OPERATING INSTRUCTIONS............................................................................ 71 6. OPERATING INSTRUCTIONS........................................................................................... 73 6.1. Overview of Operating Modes ....
Résumé du contenu de la page N° 9
Teledyne API M100E Analyzer Operation Manual Table of Contents 11.2.4. UV Lamp Shutter & PMT Offset........................................................................................................ 223 11.2.5. Optical Filters .................................................................................................................................... 224 11.3. Pneumatic Operation ......................................................................................................
Résumé du contenu de la page N° 10
Table of Contents Teledyne API M100E Analyzer Operation Manual Figure 3-12: Functional Check .........................................................................................................................41 Figure 3-13: Reporting Range Verification.......................................................................................................43 Figure 3-14: Dilution Ratio Setup ...................................................................................................
Résumé du contenu de la page N° 11
Teledyne API M100E Analyzer Operation Manual Table of Contents Figure 7-6: Rear Panel Connector Pin-Outs for COM2 in RS-485 Mode ................................................... 123 Figure 7-7: CPU Connector Pin-Outs for COM2 in RS-485 Mode.............................................................. 124 Figure 7-8: COMM – LAN / Internet Configuration Properties .................................................................... 127 Figure 7-9: COMM – Manually Configuring the Network IP
Résumé du contenu de la page N° 12
Table of Contents Teledyne API M100E Analyzer Operation Manual Figure 10-3: Simple Leak Check Fixture....................................................................................................... 217 Figure 10-4: Hydrocarbon Scrubber Leak Check Setup ............................................................................... 217 Figure 11-1: UV Absorption........................................................................................................................... 220
Résumé du contenu de la page N° 13
Teledyne API M100E Analyzer Operation Manual Table of Contents Table 5-2: Optional Pumps ...........................................................................................................................53 Table 5-3: Optional Kits.................................................................................................................................56 Table 5-4: Zero/Span Valve Operating States ..................................................................................
Résumé du contenu de la page N° 14
Table of Contents Teledyne API M100E Analyzer Operation Manual Table 12-7: Analog Output Test Function - Nominal Values ....................................................................... 273 Table 12-8: Status Outputs Check Pin Out.................................................................................................. 273 Table 12-9: Example of HVPS Power Supply Outputs ................................................................................ 277 Table 12-10: Example of
Résumé du contenu de la page N° 15
SECTION I GENERAL INFORMATION 13 04515F DCN6048
Résumé du contenu de la page N° 16
This page intentionally left blank. 14 04515F DCN6048
Résumé du contenu de la page N° 17
1. INTRODUCTION 1.1. SAFETY MESSAGES Your safety and the safety of others are very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol. These symbols are found in the manual and inside the instrument. The definition of these symbols is described below: General Safety Hazard: Refer
Résumé du contenu de la page N° 18
Introduction Teledyne API M100E Analyzer Operation Manual 1.2. M100E OVERVIEW The Model 100E (also referred to as M100E) UV Fluorescence SO Analyzer is a microprocessor controlled 2 analyzer that determines the concentration of sulfur dioxide (SO ), in a sample gas drawn through the 2 instrument’s sample chamber where it is exposed to ultraviolet light causing any SO present to fluoresce. The 2 instrument measures the amount of fluorescence to determine the amount of SO present in the sampl
Résumé du contenu de la page N° 19
Teledyne API M100E Analyzer Operation Manual Introduction Additional documentation for the M100E SO Analyzer is available from Teledyne API’s website at 2 http://www.teledyne-api.com/manuals/. APICOM software manual, part number 03945. Multidrop manual, part number 01842. iDAS Manual, part number 02837. 1.3.1. USING THIS MANUAL NOTE Throughout this manual, words printed in capital, bold letters, such as SETUP or ENTR represent messages as they appear on the analyzer’s display.
Résumé du contenu de la page N° 20
Introduction Teledyne API M100E Analyzer Operation Manual SECTION II – OPERATING INSTRUCTIONS BASIC OPERATION OF THE M100E ANALYZER Step-by-Step instructions for using the display/keyboard to set up and operate the M100E analyzer. ADVANCED FEATURES OF THE M100E ANALYZER Step-by-Step instructions for using the M100E analyzer’s more advanced features such as the iDAS system, the DIAG and VARS menus and the and the TEST channel analog output. REMOTE OPERATION OF THE M100E Analyzer Inf