Resumen del contenido incluido en la página 1
User’s Guide
Shop online at
omega.com
e-mail: info@omega.com
For latest product manuals:
omegamanual.info
HH503
Digital Thermometer
Resumen del contenido incluido en la página 2
® OMEGAnet On-Line Service Internet e-mail omega.com info@omega.com Servicing North America: U.S.A.: ISO 9001 Certified Canada: 976 Bergar OMEGA Engineering, Inc. Laval (Quebec), Canada H7L 5A1 One Omega Drive, P.O. Box 4047 Stamford, CT 06907-0047 USA Toll-Free: 1-800-826-6342 TEL: (514) 856-6928 Toll-Free: 1-800-826-6342 TEL: (203) 359-1660 FAX: (514) 856-6886 e-mail: info@omega.ca FAX: (203) 359-7700 e-mail: info@omega.com For immediate technical or application assistance: U.S.A. and Canada:
Resumen del contenido incluido en la página 3
INTRODUCTION This instrument is a 4½ digit, compact-sized portable digital thermometer designed to use external K-type and J-type thermocouples as temperature sensor. Temperature indication follows Reference Temperature/Voltage Tables (N.I.S.T. Monograph 175 Revised to ITS-90) for K-type and J-type thermo- couples. One K-type thermocouple are supplied with the thermometer. SAFETY INFORMATION It is recommended that you read the safety and operation instructions before using the thermometer. WA
Resumen del contenido incluido en la página 4
SPECIFICATIONS ELECTRICAL Temperature Scale: Celsius or Fahrenheit user-selectable Measurement Range: J-TYPE -200°C to 1050°C, (-328°F to 1922°F) K-TYPE -200°C to 1370°C, (-328°F to 2498°F) Resolution: 0.1°C or 0.2°F Accuracy: Accuracy is specified for operating temperatures over the range of 18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error. ±(0.05% rdg + 0.3°C) -50°C to 1370°C ±(0.05% rdg + 0.7°C) -50°C to -200°C ±(0.05% rdg + 0.6°F) -58°F to 2498°F ±(0.05% rdg + 1.4°F)
Resumen del contenido incluido en la página 5
ENVIRONMENTAL Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H. GENERAL Display: 4½ digit liquid crystal display (LCD) with maximum reading of 19999. Overload: "----.-" is display. Battery: Standard 9V battery. Battery Life: 100 hours typical. Auto power off: The meter key switch inactive for more than 30 minutes, press power key to resume operation. Dimensions: 192mm(H) x 91mm(W) x 52.5mm(D) Weight: 365g. 3
Resumen del contenido incluido en la página 6
OPERATING INSTRUCTIONS 1 Power Switch The key turns the thermometer on or off. In the data SET mode, can not power off, must leave data SET mode then power off. 2 °C/°F Selecting the Temperature Scale Readings are dual displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F). When the thermometer is turned on, it is set to the temperature scale that was in use when the thermometer was last turned off. To change the temperature scale, press the °C/°F key. 3 HOLD Mode (only Main
Resumen del contenido incluido en la página 7
4 Cold Junction Temperature Display mode (second display) Press C.J. TEMP key to display the thermocouple input connector cold junction temperature on second display. Press C.J. TEMP key again to exit this mode. 5 K/J T1 Input Thermocouple Type Select The K/J key switch the thermocouple input selects the K-type or J-type thermocouple as input. When the thermometer is turned on, it is set to the type selected that was in use when the thermometer was last turned off. 6 MIN MAX with Time record Mod
Resumen del contenido incluido en la página 8
The true average of all the reading taken over at least 22 hours preiod can be displayed. If 22 hours is exceeded, new averages are no longer calculated. The last calcalated value is retained as the average reading, but the actual minimum and maximum reading will continue to be captured. In the this mode, press HOLD key to stop the recording of readings, all values are frozen, press again to restart recording. To prevent accidental loss of MIN, MAX and AVG data, in this mode can only be cancelle
Resumen del contenido incluido en la página 9
8 Sec. Minu. Selecting the Time scale Reading the third displayed (Time) in either second or minute. When the thermometer turned on, it is set to second. To change the Time scale, press Sec. Minu. key. Maximum Time reading is 100 hours. If 100 hours is exceeded, reset Time to zero and restart. 9 SET mode (Relative value set, Time set and Hi/Lo Limits value set) 1. Press SET key to enter Relative valuse SET mode (Press ENTER key can escape relative valuse set mode), REL set mode. = = = =.= is dis
Resumen del contenido incluido en la página 10
4. When the thermometer is turned on. The Relative set value and Hi/Lo Limits set value that was in use when thermometer was last turned off set values. 10 Relative value Recall display mode. Press REL RECALL key to display the Relative set value on second display. Press REL RECALL key again to exit this mode. 11 Hi/Lo Limits value Recall display mode Press Hi/Lo RECALL key to display the Hi set Limit value on second display. Press Hi/Lo RECALL key again to display the Lo Limits set value on sec
Resumen del contenido incluido en la página 11
OPERATOR MAINTENANCE WARNING To avoid possible electrical shock, disconnect the thermocouple connectors from the thermometer before removing the cover. Battery Replacement Power is supplied by a 9 volt "transistor" battery. The " " appears on the LCD display when replacement is needed. To replace the battery, remove the two screws from the back of the meter and lift off the battery cover. Remove the battery from battery contacts. 9
Resumen del contenido incluido en la página 12
CALIBRATION PROCEDURE Note: The following calibration procedures should be performed only by qualified technicians who have access to the following: Equipment: The class of calibration should be 10 times greater than the measured meter. Step 1: Short the J3 with jumper. Step 2: Apply 65 mV to the T1 input from Volts Calibrator and wait till the reading becomes stable, then press the SET key until the display shows 0.1. Step 3: Apply Type K 0°C to the input T2 and wait till the reading become
Resumen del contenido incluido en la página 13
NOTES: 11
Resumen del contenido incluido en la página 14
NOTES: 12
Resumen del contenido incluido en la página 15
WARRANTY/ DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product. If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department
Resumen del contenido incluido en la página 16
Where Do I Find Everything I Need for Process Measurement and Control? OMEGA…Of Course! sm Shop online at omega.com TEMPERATURE DATA ACQUISITION Thermocouple, RTD & Thermistor Data Acquisition & Engineering Software Probes, Connectors, Panels & Assemblies Communications-Based Acquisition Wire: Thermocouple, RTD & Thermistor Systems Calibrators & Ice Point References Plug-in Cards for Apple, IBM & Recorders, Controllers & Process Compatibles Monitors Datalogging Systems