Resumen del contenido incluido en la página 1
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 1
INSTRUCTION MANUAL
Automatic Blood Pressure Monitor
Model 7100REL
ENGLISH ESPAÑOL
Resumen del contenido incluido en la página 2
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 2 TABLE OF CONTENTS Before Using the Unit Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 Operating the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 Care and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 Before Taking a Measureme
Resumen del contenido incluido en la página 3
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 3 Before Using the Unit INTRODUCTION Thank you for purchasing the ReliOn® 7100REL Automatic Blood Pressure Monitor. Fill in for future reference. DATE PURCHASED: _______________ SERIAL NUMBER: _________________ Staple your purchase receipt here Your new digital blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood’s movement through your brachial artery and converts the movements into
Resumen del contenido incluido en la página 4
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 4 SAFETY INFORMATION To assure the correct use of the product basic safety measures should always be followed including the warnings and cautions listed in this instruction manual. SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL Indicates a potentially hazardous situation which, if not avoided, could result in death or WARNING serious injury. Indicates a potentially hazardous situation which, if not avoided, may result in minor or CAUTION moderate injury
Resumen del contenido incluido en la página 5
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 5 SAFETY INFORMATION OPERATING THE DEVICE (continued) Read carefully “Taking a Measurement” No.3 of the instruction manual if your systolic pressure is known to be higher. Inflating to a higher pressure than necessary may result in bruising where the cuff is applied. Operate the device only as intended. Do not use the device for any other purpose. Dispose of the device, components and optional accessories according to applicable local regulations. Unlawfu
Resumen del contenido incluido en la página 6
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 6 SAFETY INFORMATION CARE AND MAINTENANCE Do not subject the monitor to strong shocks, such as dropping the unit on the floor. Do not submerge the device or any of the components in water. Store the device and the components in a clean, safe location. Changes or modification not approved by the manufacturer will void the user warranty. Do not disassemble or attempt to repair the unit or components. 6
Resumen del contenido incluido en la página 7
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 7 BEFORE TAKING A MEASUREMENT To ensure a reliable reading follow these recommendations: 1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes before taking a measurement. Rest for at least 15 minutes before taking the measurement. 2. Stress raises blood pressure. Avoid taking measurements during stressful times. 3. Measurements should be taken in a quiet place. 4. Remove tight-fitting clothing from your arm. 5. Sit on a chair
Resumen del contenido incluido en la página 8
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 8 Operating Instructions KNOW YOUR UNIT Main Unit: Display Battery Compartment Air Jack On/Off Start Button Components: Air Tube Four “AAA” Alkaline (LR03) Batteries INSTRUCTION MANUAL Automatic Blood Pressure Monitor Model 7100REL Air Plug (Connects to Air Jack) Arm Cuff Instruction Manual 8 ENGLISH ESPAÑOL
Resumen del contenido incluido en la página 9
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 9 UNIT DISPLAY Systolic Blood Pressure Diastolic Blood Pressure Heartbeat Symbol Memory Symbol Pulse Display Low Battery Indicator Deflation Symbol 9
Resumen del contenido incluido en la página 10
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 10 DISPLAY SYMBOLS The American Heart Association (AHA) recommends the following guideline for upper limit of normal for home blood pressure readings: AHA Home Guideline for Upper Limit of Normal BP Systolic Blood Pressure 135 mmHg Diastolic Blood Pressure 85 mmHg This is a general guideline because blood pressure depends on age and morbidity. The AHA recommends a lower home blood pressure goal for certain patients, including diabetic patients, pregnant w
Resumen del contenido incluido en la página 11
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 11 BATTERY INSTALLATION 1. Turn the main unit upside down. Battery Cover 2. Press the ▲ indicator on the battery cover and slide the cover off in the direction of the arrow. 3. Install 4 “AAA” size batteries so the + (positive) and - (negative) polarities match the polarities of the battery compartment as indicated. 4. Replace the battery cover. Slide the battery cover as indicated until it clicks into place. 11
Resumen del contenido incluido en la página 12
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 12 BATTERY INSTALLATION BATTERY REPLACEMENT Low Battery Indicator When the Low Battery Indicator appears on the display screen turn the monitor off and remove all the batteries. Replace with four new batteries at the same time. WARNING If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Contact a physician immediately. NOTE: The batteries supplied with the device may have a shorter life. CAUTION Dispose of the device, c
Resumen del contenido incluido en la página 13
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 13 APPLYING THE ARM CUFF CAUTION Read all of the information in the instruction manual and any other literature in the box before operating the unit. CAUTION This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on infants or persons who cannot express their intentions. APPLYING THE CUFF ON THE LEFT ARM 1. Make sure the air plug is securely inserted in the main unit. NOTE: If the cuff is
Resumen del contenido incluido en la página 14
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 14 APPLYING THE ARM CUFF 3. Apply the cuff to your left upper arm so the blue arrow is centered on the inside of your arm and aligned with your middle finger. The air tube runs down the inside of your arm. The bottom of the cuff should be approximately 1/2" above your elbow. 1/2 inch (1 - 2 cm) Blue Arrow Air Tube 4. Wrap the cuff firmly in place around your arm using the closure strip. 14
Resumen del contenido incluido en la página 15
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 15 APPLYING THE ARM CUFF APPLYING THE CUFF ON THE RIGHT ARM When taking a measurement using the right arm use this instruction for Step 4 on page 14. Apply the cuff to your right upper arm so the blue arrow and the air tube runs along the side of your arm. The bottom of the cuff should be approximately 1/2'' above your elbow. NOTE: Be careful not to rest your arm on the air tube. This will restrict the flow of air to the cuff. 15
Resumen del contenido incluido en la página 16
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 16 TAKING A MEASUREMENT CAUTION Operate the device only as intended. Do not use the device for any other purpose. 1. Sit on a chair with your feet flat on the floor. Place your left arm on a table so the cuff is level with your heart. Remain still and do not move your arm until the entire measurement is complete. 2. Press the ON/OFF START button. All symbols appear on the display. CAUTION Inflating to a higher pressure than necessary may result in bruis
Resumen del contenido incluido en la página 17
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 17 TAKING A MEASUREMENT 3. The cuff starts to inflate. The unit automatically inflates to 170mmHg. The cuff needs to inflate 30 to 40 mmHg higher than your expected systolic pressure. NOTES: • Do not apply more pressure than necessary. • The monitor will not inflate above 299 mmHg. • To stop the inflation or measurement, press the ON/OFF START button. The monitor will stop inflating, start deflating and will turn off. If your systolic pressure is known to
Resumen del contenido incluido en la página 18
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 18 USING THE MEMORY FUNCTION The monitor automatically stores the last measurement values in the memory. TO DISPLAY THE MEASUREMENT VALUES 1. Press and hold the ON/OFF START button until the last measurement values are displayed along with the Memory Symbol ( ). 2. Press the ON/OFF START button to turn the power off. 18
Resumen del contenido incluido en la página 19
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 19 Care and Maintenance CARE AND MAINTENANCE To keep your digital blood pressure monitor in the best condition and protect the unit from damage, follow the directions listed below: Do not forcefully bend the arm cuff or air tube. Do not fold tightly. Clean the monitor with a soft dry cloth. Do not use any abrasive or volatile cleaners. Do not attempt to wash the cuff. CAUTION Do not submerge the device or any of the components in water. Do not subject th
Resumen del contenido incluido en la página 20
7100REL_e_sp.qxd 11/4/08 3:34 PM Page 20 ERROR INDICATORS AND TROUBLESHOOTING TIPS ERROR INDICATORS SYMBOL CAUSE CORRECTION Monitor could not detect pulse wave. Take another measurement and remain still until the measurement is complete. Air plug is not connected. Cuff under-inflated. Remove the arm cuff. Cuff not applied Read “Taking a Measurement”. correctly. Take another measurement. Remove the arm cuff. Read “Taking a Measurement”. Cuff over-inflated Take another measurement without infla