2. UNPACKING THE BALANCE 2-1 Unpacking The balance is a precision instrument. Unpack the balance carefully. Keep the packing material to be used for transporting the balance in the future. The packing contents depend on the balance model. See the illustrations to confirm that everything is contained. Standard type: GP-12K/20K/30K/32K/40K/60K/61K/100K/102K Note Please confirm that the AC adapter type is correct for your
Inhaltszusammenfassung zur Seite Nr. 10
Separate display type: GP-30KS/32KS/60KS/61KS/100KS Note Please confirm that the AC adapter type is correct for your local voltage and receptacle type. 2-2 Installing the Balance Install the balance as follows: 1. Refer to “3. PRECAUTIONS” for installing the balance. 2. Adjust the leveling feet to level the balance. Confirm it using the bubble spirit level. 3. Confirm that the adapter type is correct for the local voltage
Inhaltszusammenfassung zur Seite Nr. 11
3. PRECAUTIONS To get the optimum performance from the balance and acquire accurate weighing data, note the following: 3-1 Before Use Install the balance in an environment where the temperature and humidity are not excessive. The best operating temperature is about 20°C / 68°F at about 50% relative humidity. Install the balance where it is not exposed to direct sunlight and it is not affected by heaters or air conditioners. Install the balance where it is free of dust. Install
Inhaltszusammenfassung zur Seite Nr. 12
This balance uses a strong magnet as part of the balance assembly, so please use caution when weighing magnetic materials such as iron. If there is a problem, use the underhook on the bottom of the balance to suspend the material away from the influence of the magnet. Cancel the temperature difference between a sample and the environment. When a sample is warmer (cooler) than the ambient temperature, the sample will be lighter (heavier) than the true weight. This error is due to a
Inhaltszusammenfassung zur Seite Nr. 13
3-5 Display Symbols and Key Operation Key operation Key operation affects how the balance functions. The basic key operations are: “Press and release the key immediately” or “Press the key” = normal key operation during measurement “Press and hold the key” Display symbols Each key, when pressed or when pressed and held, functions as follows: Key When pressed When pressed and held Turns the display ON and OFF. The standby indicator is di
Inhaltszusammenfassung zur Seite Nr. 14
3-6 Smart Range Function The GP-32K, GP-32KS and GP-102K are equipped with two ranges. The precision range has a higher resolution. The standard range has normal resolution. The range is switched automatically, depending on the value displayed. Pressing the RE-ZERO key allows weighing in the precision range, regardless of the tare value. (Smart range function) The range can be fixed to the standard range, by pressing the SAMPLE key. Note Once the range is switched to the standard range,
Inhaltszusammenfassung zur Seite Nr. 15
4. WEIGHING UNITS 4-1 Units With the GP series balance, the following weighing units and weighing modes are available : A unit or mode can be selected and stored in the function table as described in “4-2 Changing the Units”. If a weighing mode (or unit of weight) has been turned off, that mode or unit will be missing in the sequence. Tael has four varieties, one of which can be selected and installed at the factory. For details about the units and modes, see the tabl
Inhaltszusammenfassung zur Seite Nr. 16
The tables below indicate the weighing capacity and the minimum display for each unit, depending on the balance model. Unit GP-12K GP-20K GP-30K GP-61K GP-40K Minimum Minimum GP-30KS GP-61KS display display Capacity Capacity Gram 12000.0 21000.031000.061000.0 0.141000.0 0.5 Kilogram 12.0000 21.000031.0000 61.0000 0.0001 41.0000 0.0005 Ounce (Avoir) 423.290 740.755 1093.495 2151.710 0.005 1446.24 0.02 Pound 26.4555 46.297068.3435 134.4820 0.0005 90.390 0.002 Pound/Ounce 26L 7.29 46L
Inhaltszusammenfassung zur Seite Nr. 17
Unit GP-32K/ GP-32KS Standard range Precision range Capacity Minimum display Capacity Minimum display Gram 31000 16100.0 0.1 Kilogram 31.000 0.0016.1000 0.0001 Ounce (Avoir) 1093.50 0.05 215.170 0.005 Pound 68.345 0.00513.4480 0.0005 Pound/Ounce 68L 5.5 0.1 13L 7.17 0.01 Troy Ounce 996.65 0.05 196.120 0.005 Metric Carat 155000 5 30500.0 0.5 Mom me 8266.5 0.5 1626.65 0.05 Pennyweight 19933 1 3922.4 0.1 Tael (HK general, 820.10 0.05161.380 0.005 Singapore) Tael (HK jewe
Inhaltszusammenfassung zur Seite Nr. 18
4-2 Changing the Units The units or modes can be selected and stored in the function table. The sequence of displaying the units or modes can be arranged so as to fit the frequency of use in the function table. Select a unit or mode and arrange the sequence of display as follows: 1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the key. 2 Press the SAMPLE key several times to display Unit . 3 Press the PRINT key to enter the unit selecti
Inhaltszusammenfassung zur Seite Nr. 19
5. WEIGHING 5-1 Basic Operation (Kilogram Mode) 1 Place a container on the weighing pan, if necessary. Press the RE-ZERO key to cancel the weight (tare). The balance displays 0.0000 kg. (The decimal point position depends on the balance model.) 2 Place a sample on the pan or in the container. 3 Wait for the stabilization indicator to be displayed. Read the value. 4 Remove the sample and container from the pan. Notes To use another unit, press the MODE key and select an appropria
Inhaltszusammenfassung zur Seite Nr. 20
5-2 Counting Mode (PCS) This is the mode to determine the number of objects in a sample based on the standard sample unit mass. Unit mass means the mass of one sample. The smaller the variables in each sample unit mass is, the more accurate the counting will be. The GP series balance is equipped with the Automatic Counting Accuracy Improvement (ACAI) function to improve the counting accuracy. Note If the sample unit mass variable, the difference from sample to sample, is too large, it ma
Inhaltszusammenfassung zur Seite Nr. 1
GP SERIES
GP-12K/GP-20K /GP-30K/GP-32K
GP-40K/GP-60K/GP-61K
GP-100K/GP-102K
GP-30KS/GP-32KS
GP-60KS/ GP-61KS/GP-100KS
Precision Balance
INSTRUCTION MANUAL
WM+PD4000273D
Inhaltszusammenfassung zur Seite Nr. 2
© 2006 A&D Company Ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of A&D Company Ltd. The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice.
Inhaltszusammenfassung zur Seite Nr. 3
CONTENTS Basic operation 1. INTRODUCTION ............................................................................................................................ 3 1-1 About This Manual ..............................................................................................................................3 1-2 Features ..............................................................................................................................................3 1-3 Compliance ...
Inhaltszusammenfassung zur Seite Nr. 4
9-8 Data Format Examples .....................................................................................................................46 9-9 Clock and Calendar Function............................................................................................................48 9-10 Comparator Function ......................................................................................................................50 9-11 Adding the Comparison Results............................
Inhaltszusammenfassung zur Seite Nr. 5
1. INTRODUCTION This manual describes how the GP series balance works and how to get the most out of it in terms of performance. Read this manual thoroughly before using the balance and keep it at hand for future reference. 1-1 About This Manual This manual consists of the following five parts: Basic operation..............................Describes precautions on handling the balance, balance construction and basic balance operation. Adapting to the environment..........Describes r
Inhaltszusammenfassung zur Seite Nr. 6
Hold Function, provided for weighing a moving object such as an animal. Multiple Weighing Units, with most of the common units used around the world. Density Mode, for calculating the density of a solid. Accumulation Function, adding the weight values and outputting the sum. Adjustable Display Position, the standard type balance has a swing-arm to adjust the display position for optimum viewing. Separate display type balances (GP-30KS/32KS/60KS/61KS/100KS) are also available.
Inhaltszusammenfassung zur Seite Nr. 7
1-3 Compliance Compliance with FCC Rules Please note that this equipment generates, uses and can radiate radio frequency energy. This equipment has been tested and has been found to comply with the limits of a Class A computing device pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to provide reasonable protection against interference when equipment is operated in a commercial environment. If this unit is operated in a residential area, it may cause some interfe
Inhaltszusammenfassung zur Seite Nr. 8
6