ملخص المحتوى في الصفحة رقم 1
MANUAL ADDENDUM
ULTRA SENSITIVITY MODEL
M200EU
NO/NO /NOx ANALYZER
2
(Addendum to M200E Manual PN 04410)
Teledyne Advanced Pollution Instrumentation
9480 Carroll Park Drive
San Diego, CA 92121-5201
USA
Toll-free Phone: 800-324-5190
Phone: 858-657-9800
Fax: 858-657-9816
Email: api-sales@teledyne.com
Website: http://www.teledyne-api.com/
05385 Rev C
Copyright 2009-2010 DCN 5731
Teledyne Advanced Pollution Instrumentation 14 May 2010
ملخص المحتوى في الصفحة رقم 2
ملخص المحتوى في الصفحة رقم 3
Ultra Sensitivity Model M200EU NO/NO /NOx Analyzer Addendum to M200E 2 TABLE OF CONTENTS 1. Overview.................................................................................................................... 1 2. Specifications, Agency Approvals, and Warranty..................................................... 2 3. Pneumatics................................................................................................................. 4 4. Spare Parts List...................
ملخص المحتوى في الصفحة رقم 4
ملخص المحتوى في الصفحة رقم 5
Ultra Sensitivity Model M200EU NO/NO /NOx Analyzer Addendum to M200E 2 1. OVERVIEW The model M200EU is a close derivative of the Model 200E NOx Analyzer. The main differences are as follows. The M200EU is more sensitive at low NOx levels than a standard 200E. The instrument has a user selectable full scale range of 0-5 PPB to 0-2000 PPB of NOx. The reaction cell is gold plated. In addition, there is a pre-reactor, which allows the measurements of any hydrocarbon interferents in t
ملخص المحتوى في الصفحة رقم 6
Ultra Sensitivity Model M200EU NO/NO /NOx Analyzer Addendum to M200E 2 2. SPECIFICATIONS, AGENCY APPROVALS, AND WARRANTY Min/Max Range Min: 0-5 ppb of Full Scale (Physical Analog Output) Max 0-2000 ppb of Full Scale 3 Measurement Units ppb, ug/m (user selectable Single, independent or auto ranges) 1 Noise at Zero <25 ppt RMS 1 Noise at Span <0.25% RMS of reading above 50 ppb 1 Lower Detectable Limit 50 ppt RMS <0.1 ppb / 24 hours 2 Zero Drift <0.2 ppb / 7 days <0.5% of reading or
ملخص المحتوى في الصفحة رقم 7
Ultra Sensitivity Model M200EU NO/NO /NOx Analyzer Addendum to M200E 2 COM 1: RS-232; Serial I/O COM 2: RS-232 or RS-485 (Ethernet interface available) Baud Rate: 300-115200 4 Environmental Installation Category (Over-voltage Category) II, Pollution Degree 2 US EPA: Reference Method Number RFNA 1194-099 Certifications CE: EN61326 (1997 w/A1:98) Class A, FCC Part 15 Subpart B Section 15.107 Class A, ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 &A2;97) Class A (1) As define
ملخص المحتوى في الصفحة رقم 8
Ultra Sensitivity Model M200EU NO/NO /NOx Analyzer Addendum to M200E 2 3. PNEUMATICS Refer to the following pneumatic diagram. It shows the internal pneumatics of the M200EU in its standard configuration, along with the optional Zero/Span valves. The sample flow is about 1000 cc/min. Also during Auto Zero mode, the sample gas is routed through a pre-reactor before the mixture goes through the reaction cell. Since the reaction rate constant between NO and ozone is several time faster than
ملخص المحتوى في الصفحة رقم 9
Ultra Sensitivity Model M200EU NO/NO /NOx Analyzer Addendum to M200E 2 4. SPARE PARTS LIST Note: Only parts different from M200E are listed below. Refer to the M200E users manual for the common spare parts. Part Number Description 011340800 Assy, Sensor, M200EU 024600100 Assy, Prereactor NO/NOx valve, M200EU 011420700 Assy, NOx cell, Gold plated, M200EU 011440200 Assy, PMT, M200AU/EU 041800600 PCA, PMT Preamp, VR, M200EU 001270001 Manifold, Gold Plated, NOx 001330001 Sleeve, Gold Pl