ملخص المحتوى في الصفحة رقم 1
User’s Guide
http://www.omega.com
e-mail: info@omega.com
DPG5000
PRESSURE GAUGES
M3366/0606
ملخص المحتوى في الصفحة رقم 2
Where Do I Find Everything I Need for Process Measurement and Control? OMEGA...Of Course! DATA ACQUISITION TEMPERATURE ✓ Data Acquisition& Engineering ✓ Thermocouple, RTD & Thermistor Software Probes,Connectors, Panels & ✓ Communications-Based Assemblies AcquisitionSystems ✓ Wire: Thermocouple, RTD& ✓ Plug-inCards for Apple, IBM& Thermistor Compatibles ✓ Calibrators& IcePoint References ✓ DataloggingSystems ✓ Recorders,Controllers &Process
ملخص المحتوى في الصفحة رقم 3
M3366/0606 DPG5000 Table of Contents Warranty/Disclaimer.........................................................................2 Return.Requests/Inquiries.................................................................3 Instructions.......................................................................................4 . Zero.Trimming................................................................................4 . Use.of.Auto-Tare.Feature................................................
ملخص المحتوى في الصفحة رقم 4
DPG5000 M3366/0606 • Warranty/Disclaimer OMEGA ENGINEERING, INC.warrants thisunit tobe free of defects in materials and workmanshipfor a periodof 13 months from date of purchase.OMEGA ’s Warranty addsan additional one (1)month grace period to thenormal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’scustomers receive maximumcoverage on each product. If theunit malfunctions, it mustbe returned
ملخص المحتوى في الصفحة رقم 5
M3366/0606 DPG5000 • Return Requests/Inquiries Direct allwarranty andrepair requests/inquiriesto the OMEGA Customer ServiceDepartment. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR)NUMBER FROMOMEGA ’SCUST OMER SERVICE DEPARTMENT (INORDER TO AVOID PROCESSING DELAYS). The assigned ARnumber shouldthen bemarked on the outside ofthe return packageand on any correspondence. The purchaser is r
ملخص المحتوى في الصفحة رقم 6
DPG5000 M3366/0606 • Instructions NOTE:.All.units.are.factory.calibrated.prior.to.shipment. 1. Zero Trimming NOTE:.Some.units.are.equipped.with.zero.“knob”. If itbecomes necessary tore-adjust the “Zero” on the display, this can be accomplished by turning the trimpot marked “Z” just tothe leftof theOn/Of f push button.On the DPG5000there is either a zero adjustment knob ora black nyloncover screw . If the gauge has a cover screw, it is necessary to remov
ملخص المحتوى في الصفحة رقم 7
M3366/0606 DPG5000 4. Re-Calibration This procedure requiresa knownpressure sourceof at least ±0.1% accuracy in order to fullyutilize the accuracy potential of the DPG5000 (If notavailable, the gauge can be returned to OMEGA for re-calibration.) Procedure: A. Ensure the Gaugeis at0 psig (orvacuum if absolute), and adjust the zero asper instructions in step 1. B. Applyfull scale pressureto the pressure port and adjust thespan (“S”) pot unt
ملخص المحتوى في الصفحة رقم 8
DPG5000 M3366/0606 7. 4-20 mA Transmitter Option When equipped with thisoption thegauge no longer operates frombatteries butinstead is“Loop Powered”. Theloop connection is madeto a 6 pin receptacle located at therear of the unit (a matingplug is supplied). A voltageof between 9 and32 VDC mustbe maintained at this connection (Pin Ais “+” and Pin B is“-” see sketch) to insureproper operation. Completion ofthe earth or system
ملخص المحتوى في الصفحة رقم 9
M3366/0606 DPG5000 8. External Power (24 VDC) and 0 to 5 VDC Analog Output Option When equipped with this option thegauge no longer operates frombatteries but instead is“externally powered”.Both the 24* VDCinput power and the 0to 5 VDC Analog Output areconnected to the 6 pincircular connectoron the rear of the unit. Pin connections areas follows: *Actual.range.of.input.voltage.is.12.to.31.VDC. 24 VDC External Power and 0-5 V Analog Output PIN -A- PL
ملخص المحتوى في الصفحة رقم 10
DPG5000 M3366/0606 Notes
ملخص المحتوى في الصفحة رقم 11
Thank You for Purchasing OMEGA Products OMEGA’s policy isto make running changes, not modelchanges, whenever an improvement is possible. Thisaf fordsour customers the latest in technology and engineering. Itis thepolicy of OMEGA tocomply with all worldwide safety andEMC/ EMIregulations that apply. OMEGA isconstantly pursuing certification of its products tothe European New ApproachDirectives. OMEGA willadd the CEmark to every appr
ملخص المحتوى في الصفحة رقم 12
® OMEGAnet On-Line Service Internet e-mail www.omega.com info@omega.com Servicing North America: USA: One Omega Drive, P.O.Box 4047 ISO.9001.Certified Stamford, CT 06907-0047 TEL: (203) 359-1660 FAX: (203)359-7700 e-mail: info@omega.com Canada: 976Ber gar Laval (Quebec) H7L 5A1 TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail:info@omega.ca For immediate technical or application assistance: ® USA and Canada: Sales Service: 1-800-826-6342/ 1-800-TC-OMEGA