ملخص المحتوى في الصفحة رقم 1
Milwaukee
USER MANUAL
PORTABLE
DISSOLVED OXYGEN METER
MODEL: MW600
Smart DO Meter
ملخص المحتوى في الصفحة رقم 2
PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the meter. Install the battery into the battery clip connector while observing polarity. D.O. probes are shipped dry. Assemble the probe and prepare it for use, connect it to the meter and proceed as follows. 1. Remove the red and black plastic cap. This cap is used for shipping purposes only and can be thrown away. 2. Wet the sensor by soaking the bottom 2½ cm of the probe in elect
ملخص المحتوى في الصفحة رقم 3
PROBE POLARIZATION: • A dissolved oxygen probe must be polarized to function properly. • To polarize the probe, the fully assambled probe must be connected to the meter and the meter must be on. • During polarization (and during measurement), approximately 800 mV is applied to the cathode and anode inside the membrane and a chemical reaction occurs. During the polarization period excess oxygen in the electrolyte is consumed. During this phase, probe movement that “moves” the electrolyte will yie
ملخص المحتوى في الصفحة رقم 4
Zero Calibration: • Dip the probe into MA9070 zero oxygen solution and stir gently for 2-3 minutes. • Wait for the probe to reach it’s lowest stable reading. • Adjust the zero D.O. calibration trimmer (located on the front panel) until the display reads "0.0". Slope Calibration: It is suggested to perform the slope calibration in water saturated air. • Rinse the probe with a large amount of clean water to remove any residual zero oxygen solution. • Dry the probe tip and allow a few minutes for t
ملخص المحتوى في الصفحة رقم 5
TAKING MEASUREMENTS: Verify the probe is polarized and the probe and meter have been calibrated. Remove the protective cap from probe. Immerse the tip of the probe in the sample to be tested. For accurate dissolved oxygen measurements a minimum water movement of 0.3 m/sec is required. This is to ensure that the oxygen-depleted membrane surface is constantly replenished. A moving stream will provide adequate circulation. To quickly check if the water speed is sufficient, wait for the reading to s
ملخص المحتوى في الصفحة رقم 6
ALTITUDE COMPENSATION: The displayed measurements are referenced to sea level pressures. At higher elevations, oxygen solubility decreases (thus at higher elevations actual oxygen concentrations are really lower than the displayed value). The table below illustrates the changes in the solubility of oxygen in air saturated fresh water as a result of changes in elevation. The table can also be used to correct the displayed measurement. If the meter was calibrated at an elevation above sea level,
ملخص المحتوى في الصفحة رقم 7
SALINITY COMPENSATION: The table below illustrates the change in the solubility of oxygen in air saturated water as a result of chloride concentration or salinity. The table can also be used to correct the displayed measurement. If you are making measurements in salt water and know the chloride concentration (or salinity), you can multiply your reading by the ratio of (ppm at the chloride concentration) / (the ppm at 0 g/L chloride) at the temperature of measurement to compensate for the salt e
ملخص المحتوى في الصفحة رقم 8
For example, if the measurement displayed at 10°C is 5 ppm, but the sample has 20 g/L of chloride, to correct your measurement multiply the displayed measurement by the ratio of (ppm reading at 20 g/L) / (ppm reading at 0 g/L) = 5.0 ppm X (8.9 ppm/11.3 ppm) = 3.93 ppm (or 3.9 ppm Chloride or Salinity corrected). PROBE & MEMBRANE MAINTENANCE: The D.O. probe body is made of reinforced plastic for maximum durability. A thermistor temperature sensor provides temperature measurements of the tested sa
ملخص المحتوى في الصفحة رقم 9
The Platinum cathode should always be bright and untarnished. If it is tarnished or stained, due to contact with certain gases or a damaged membrane cap, the cathode should be cleaned. You can use a clean lint-free cardboard or cloth. Rub the cathode very gently side to side 4-5 times. This will be enough to polish and remove any stains without damaging the platinum tip. Rinse the probe with deionized or distilled water and install a new membrane cap using fresh electrolyte (see page 1). Recalib
ملخص المحتوى في الصفحة رقم 10
BATTERY REPLACEMENT: When the battery becomes weak the meter will display the low battery indicator " ". When this appears, only a few hours of battery life remain. A low battery will result in unreliable measurements. Prompt battery replacement is required. Battery replacement must take place in a non-hazardous area using an alkaline 9V battery. Turn the meter off, slide the battery compartment cover located at the rear of the meter off and replace the 9V battery with a new one. Make sure the b
ملخص المحتوى في الصفحة رقم 11
SPECIFICATIONS: MW600 RANGE 0.0 to 19.9 mg/L RESOLUTION 0.1 mg/L ACCURACY (@25°C) ±1.5% Full Scale TEMPERATURE Automatic from 0 to 30°C COMPENSATION CALIBRATION Manual on 2 points (zero and slope) LCD 3½ digits with symbols PROBE MA840 (included) ENVIRONMENT 0 to 50°C, 95% RH max. BATTERY TYPE 9V alkaline (included) BATTERY LIFE approximately 70 hours of use DIMENSIONS 143 x 80 x 32 mm WEIGHT 220 g (with battery) meter only WARRANTY: This instrument is warranted from all defects in materials and
ملخص المحتوى في الصفحة رقم 12
ISTMW600 01/10 Milwaukee Authorized Dealer: